This Technical Report describes the general principle and technical details of the methods of measurement of the DC magnetic properties of permanent magnet materials in an open magnetic circuit using a superconducting magnet (SCM).
This method is applicable to permanent magnet materials, such as those specified in IEC 60404-8-1, the properties of which are presumed homogeneous throughout their volume.
There are two methods:
– the SCM-Vibrating Sample Magnetometer (VSM) method;
– the SCM-Extraction method.
This document also describes methods to correct the influence of the self-demagnetizing field in the test specimen on the demagnetization curve measured in an open magnetic circuit. The magnetic properties are determined from the corrected demagnetization curve.
NOTE These SCM-methods can determine the magnetic properties of permanent magnet materials with coercivity higher than 2 MA/m. The methods of measurement in a closed magnetic circuit specified in IEC 60404-5 can lead to significant measurement error due to saturation effects in the pole pieces of yoke for the magnetic materials with coercivity higher than 1,6 MA/m (see IEC 60404-5).